compared three body temperatures. The rectal and auxiliary temperatures were measured with glass mercury thermometers. The aural temperature was measured by an infrared thermometer. They concluded tha


compared three body temperatures. The rectal and auxiliary temperatures were measured with glass mercury thermometers. The aural temperature was measured by an infrared thermometer. They concluded that the infrared tympanic thermometer could be applied in an emergency room setting. Rosenthal and Leslie compared the accuracy of an electronic and an infrared thermometer with traditional glass mercury thermometry in the 95% confidence level. They found the average difference between the reading values of infrared thermometer and glass thermometer was within 0.1 °C. Stavem et al.  assessed the accuracy of infrared ear thermometry, by measuring the rectal and esophageal temperature with thermistor thermometer and ear temperature by infrared thermometer. The mean of two ear temperatures had better agreement with the rectal temperatures.
The inconsistency of the measurement may be explained by the performance, confidence level, and uncertainty of the infrared thermometer. The factors affecting the performance of infrared tympanic thermometer were discussed by Heusch et al.  Their results indicated that the handedness, sex and age were the significant factors conflicted the accuracy of the ear temperature measurement. Pusnik and Drnovsek found that several factors could affect the performance of infrared ear thermometers; such as the position of thermometer related to the aperture of the blackbody, the drift of a thermometer due to heating, the amount of times probe covers were used and the differences of probe covers.
The calibration of the infrared thermometer is very important to ensure its performance. Because the emissivity of canal is very close to unity, the temperature of a black body cavity is usually served as the standard temperature for calibration. Cascetta developed a blackbody cavity that consisted of a copper cylinder. This cylinder contacted with the copper plate and maintained a constant temperature by circulating water provided by a hot water bath, Pusnik et al. compared the measurements of infrared thermometer performed at several blackbodies. Three cavity shapes were found that could be served as suitable standards for IR calibrations. Simpson et al. described a commercial ear thermometer calibrator that could be traceable to ITS-90.
Pusnik et al. defined some important terms for the measurement of the infrared thermometer. The performance of accuracy is closeness of the measurement result of thermometer and the true value of a measurement. The difference between the average value of several measurements and the true value was called systematic error. The persity of inpidual measurements is presented as random errors. However, the variability of measurements was not the only source that induced from the difference between one measured values and average values of several measurements. Other components could produce a variability source for measurement. The uncertainty of measurements have been defined and explained by the guide to the expression of uncertainty. The components of uncertainty of an infrared ear thermometer were listed . These components included the repeatability of an IR thermometer, the reference thermometer, the blackbody radiator, the transducer of instrument and the resolution of the IR thermometer.
Recently, uncertainty evaluation had been widely applied for physical and chemical sensors. The effects of calibration equation on the measurement performance have been studied. It is very useful to study the influence of the factors affecting the performance of the IR thermometer. As far as the authors know, there have been no reports of uncertainty evaluations of an infrared tympanic thermometer.
The objectives of this study are to evaluate the accuracy and to calculate the uncertainty of two types of infrared tympanic thermometer according to ISO GUM (Guide to the expression of uncertainty in measurement). An adequate calibration equation is first established. Then the effect of the calibration equation on the accuracy and uncertainty was compared. 红外鼓膜温度计英文文献和中文翻译(3):http://www.youerw.com/fanyi/lunwen_16489.html